Portable Anesthesia Machine for MRI, Model 2200  
Model 2200 MRI Portable Anesthesia Machine is fully Portable and low maintenance This compact, robust flexible machine is easy to move. It is pneumatically powered and contains an electronic ventilation monitor that is A/C or battery powered. It may be mounted on a table or mobile cart.

Model 2200 MRI Portable Anesthesia Machine is a simple to operate unit containing all of the modes and necessary components to deliver a basic anesthetic, including a mechanical ventilator and a single agent vaporizer. All of the safety systems required are employed onboard and easy to utilize. Physiologic monitoring of the operators choice, may be employed as a separate, but necessary device.

Item Number: Description:
03-2200 Portable Anesthesia Machine for MRI, Model 2200
SPECIFICATIONS

 
 
 
Height 15 inches
Depth 8 inches
Width 21 inches
Weight  
Free-Standing 45 LBS
Fully Enclosed 60 LBS
Case Material Mil Spec ABS Plastic
All other Materials Aluminum, brass and plastic
Operating Temperature Range 35 to 95 degrees, F.
Gas Supply Sources 45-55 PSI Oxygen and Air
Fresh Gas Flow >30 Liters per Minute
Oxygen Flush Valve Recessed, self-closing, push button color coded and labeled, provides 45-70 L/min. flow while depressed
CO2 Absorber System King Systems KAB-1, prefilled or
refillable models
C02 Absorber Capacity 400 grams soda-lime
Absorber Manifold Mounts to main frame
Directional Valves Built-in the absorber
Bag-Ventilator Switch Hand-operated selector
Bellows Latex free, upward inflating, range from
0 to 1.6L
Common Gas Outlet Size indexed, quick connect
Tubing Circuit King Systems F-360-61 universal “F” or any regular anesthesia circle system
Gas Pressure Hoses DISS/thread indexed connectors
Gas Inlet Manifold DISS/thread indexed, one way valves
Oxygen and Air Supply Gauges 0-3000 PSI reading color coded and
clearly labeled
Oxygen Supply Failure Alarm Pneumatically actuated when oxygen pressure falls below 30 PSIG
Oxygen and Air Flowmeters Calibrated and scaled 0-16LPM color coded
and labeled clearly
Oxygen Analyzer Fuel cell type sensor
Auxiliary Oxygen Flowmeter Scaled 0-15LPM color coded and clearly
labeled
Vaporizer Single device, bolt-mounted, Penlon
Delta-Sigma, temperature compensated
choice of agent
Patient Airway Pressure Gauge Scaled–10 to +125 cmH2O
Total System Leakage @ 30cm H2O 0ML/min
@ 80cm H2O 160ML/min
Internal System Compliance @ 20cm H2O 1.1ML/cm H2O
@ 40cm H2O 1.3ML/cm H2O
Internal System Resistance @ 1.0L/sec gas flow 4.11cm H2O
@ 0.5L/sec gas flow 1.80cm H2O
APL Valve Pressure Drop @ 3.0L/min gas flow 0.12cm H2O
@ 30L/min gas flow 1.03cm H2O
Adjustable Pressure-Limit Adjustable, 0 to 40 cmH2O
Valve (APL)  
Ventilator-Bellows High Pressure Preset to 70 cmH2O
Relief Valve  
Mechanical Ventilator Pneumatically operated, drives bellows only
Volume Range 0-1.6L
Respiratory Rate Range 0-99 Breaths per Minute
Inspiratory Time Range .2 to 2.0 Seconds


 

RESPIRATORY MONITOR/ALARM SPECIFICATIONS

 
 
 
Flow Sensor Specifications  
Flow Range (L/min)  
Neonatal Flow
 
0.25 to 25
Pediatric/Adult Flow 2 to 180
Flow Accuracy
Neonatal Flow
 
Greater of ±3% of reading or 0.125 L/min
Pediatric/Adult Flow Greater of ±3% of reading or 0.5 L/min
Minute Volume Range (L/min)  
Neonatal Flow 0.1 to 15
Pediatric/Adult Flow 2 to 60
Tidal Volume Range  
Neonatal Flow 1 to 100
Pediatric/Adult Flow 200 to 3000
Airway Pressure Range (cm H2O)  
Neonatal Flow -120 to 120
 Pediatric/Adult Flow -120 to 120
Added Dead Space (ml)  
Neonatal Flow Less than 1ml
Pediatric/Adult Flow 6.5
Operating Temperature 5 to 45°C
Storage Temperature -30 to +65°C
Humidity 2 to 95% relative humidity, non-condensing
Barometric Pressure 400 to 800mmHg

Note
:
Accuracy applies to the given range. Actual display range may be wider.
“Internal Flow” refers to the basis for the flow-drived parameter calculations.
The accuracy specification is for room air at a nominal temperature of 25°C and
a barometirc pressure of 760mmHg. Flow sensor inlet conditions are any of the
endotracheal tube sizes 3,5 and 7mm positioned to the natural curve of the tube.
The inspiratory inlet condition is several inches of straight breathing circuit tubing.

 
FiO2 Parameter  
Probe Galvanic fuel cell
FiO2 Range 0 to 100%
FiO2 Resolution 1%
FiO2 Accuracy ± 2% if calibrated with 100% O2
FiO2 Response < 10 seconds to 90%
Life of Oxygen Sensor 750,000 O2-percent hours (20 months in continuous 50% O2)
Operating Airway Pressure Range ± 200 mmHg from ambient